- Datex Ohmeda S/5 Monitor User Manual
- Datex Ohmeda S 5 Monitor
- Datex Ohmeda S 5 Monitor User Manual Free
Datex-Ohmeda S/5 Light Monitor, F-LM1, F-LMP1 is intended for stationary and mobile monitoring of patient’s ECG, SpO2, body temperature, invasive and non-invasive blood pressures as well as respiration. S/5 Light Monitor User's Reference Manual The User’s Reference Manual provides detailed instructions and references for the operation. 1 Document no. M1031517-02 About this manual Notes to the reader This Technical Reference Manual is intended for service personnel and engineers who will install, service and maintain Datex-Ohmeda S/5 Compact Anesthesia Monitors running the software license L-CANE05 or L-CANE05A and S/5 Compact Critical Care Monitors running the software license L-CICU05 or L-CICU05A.
One moment please..
- Parts
- Videos
- News
DESCRIPTION
S/5TM Compact Anesthesia MonitorA complete monitor for anesthesia, the Datex-Ohmeda S/5 anesthesia monitor is a multi-parameter monitor. The Datex-Ohmeda S/5 anesthesia monitor features a large screen for easy viewing in the operating room. It is an ideal monitor for any setting, including the hospital, surgery center, small OR and recovery. The Datex-Ohmeda S/5 anesthesia monitor is a top of the line unit from the leader in anesthesia monitoring.
FORUMSView All (3)
Ask a New Question2Repliesa year ago | a year agoBoot loop? Hello,I've recently encountered a problem that boot looped my monitor so I decided to change de the power management board for a new one and this resolved the issue.. for 3 weeks or so.Now Im facing the same problem but it doesn't go into a boot loop everytime I boot it as it used to do a few weeks ago.. It can take up to 10 times before it does.. It makes the monitor unreliable which means I can't send it back to service.What would you do?Thnkas in advance!Reply |
-Peterann 4 years ago | 4 years agoS/5 Hemodynamic Module; E-PRESTN-00; Multiparameter Module Hi, This module initially shows all parameters on screen, but soon blank screen with message disconnection. If any one had similar problems? Thanks for your advises in advance! Reply |
2Replies5 years ago | 5 years agoS/5 Compact Manual Does anyone have a GE Datex S/5 compact users manual? Reply |
DOCUMENTS / MANUALSView All
FEATURES
- User interface specifically designed for the critical care environment
- Integrates basic hemodynamic and all airway gas measurements
- Unique capabilities of patient spirometry, continuous gas exchange/metabolics, automated tonometry and neuromuscular transmission
- Measures ECG, IBP, NIBP, SpO2, Temp, and EtCO2
- A compact 2+2 slot monitor using standard S/5 Datex-Ohmeda modules
- Provides unique capabilities of patient spirometry, continuous gas exchange/metabolic, automated tonometry and neuromuscular transmission
- Large 12.1' color display
- Color Screen
- ECG
- NiBP - Non-Invasive Blood Pressure
- SpO2 (pulse oximetry)
- EtCO2
- 5 Agent Gas analysis to monitor Sevoflurane & Desflurane agents
- Printer
SPECIFICATIONS
Height | 13.4 in |
Length | 8.1 in |
Weight | 24.2 lbs |
Width | 13.2 in |
One moment please..
- Parts
- Videos
- News
DESCRIPTION
Datex Ohmeda S/5 Monitor User Manual
A lightweight and fully featured monitor
The Datex AS/3 Multi-Parameter Patient Monitor is a light weight and fully featured monitor with three lead ECG, Temp, Sp02, and niBP capabilities.
FORUMSView All (21)
Ask a New Question4Repliesa year ago | a year agoDatex Multiparameter module - NIBP leaking question plus need silicone tubing size Hello and thanks for reading. We have a Datex / Ohmeda AS/3 with multiparameter module. The NIBP 'works' but I'm suspecting a leak somewhere in the system as it seems have to 'hunt' for the correct pressure when doing a reading and the observed mmHg on the screen descends too quickly when reading NIBP. Many times it just reports a leak. We have ruled out the tubing and cuff (replaced plus they work just fine on another monitor).Internal inspection of the multifunction module that slides into the AS/3 reveals no obvious defect. I poked around on the silicone tubing that goes from the connector to the pump / transducer and removed it and it seemed a bit loose. In addition I slightly tore one end of the tubing. It's precisely fitted, so cutting off the torn 5mm yields a tube that is too short to fit back in place. This tubing is 3.5mm ID and 7mm OD. Can't find anything that exactly fits this tubing size but I'm about to order a few that are close. So, two question:1) Do you have a description for this exact tubing?2) What other parts may I look at in the NIPB that might 'leak' (pump, transducer, solenoid)? I can get a working module for $200 but I hate to trash this [email protected]Reply |
-nefron82 3 years ago | 3 years agoAs/3 doesn't starting.. I have Datex Ohmeda as/3 monitor on my Datex as/3 ADU. Now when I try to turn it on, it doesn't work. Was black display. I change a cpu board and change battery in Power unit. Now it starting for 15 sec, I see that monitor want to start but after information massage it always restarting.Help me please.Reply |
1Reply3 years ago | 3 years agoAS/3 compact display off after standby Have an AS/3 compact unit.. works fine when first plugged in. If you then use the switch to put it into standby, the LED flashes as expected. If you then hit the switch again to take the unit out of standby the unit seems to power back up (makes all the normal noises, etc) but the display does not come back on. The only way to get the display back seems to be to unplug the power cord and plug back in again. Troubleshooting suggestions? Cheers, Mike Reply |
DOCUMENTS / MANUALSView All
Datex Ohmeda S 5 Monitor
FEATURES
Datex Ohmeda S 5 Monitor User Manual Free
- ECG, Sp02, Temp., NIBP. And BP patient monitor
- 2 hour battery life with swappable battery bays
- 12.4 x 16.1 x 5.3 inches, small and compact
- 14” color video CRT display
- Anesthesia alarm system classified into three categories